EMC guidance and manufacturer's declaration
Guidance and manufacturer's declaration-electromagnetic emissions
The MA801f is intended for use in the electromagnetic environment specified below. The customer or the
user of the MA801f should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
Group 1
The MA801f uses RF energy only for its internal function.
CISPR 11
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
Class B
The MA801f is suitable for use in all establishments,
CISPR 11
including domestic establishments and those directly
connected to the public low-voltage power supply network
Harmonic
Not applicable
that supplies buildings used for domestic purposes.
emissions IEC
61000-3-2
Voltage
Not applicable
fluctuations/flicker
emissions IEC
61000-3-3
16
Guidance and manufacturer's declaration-electromagnetic immunity. The MA801f is intended for use in the
electromagnetic environment specified below. The customer or the user of the MA801f should assure that
it is used in such an environment.
Immunity test
IEC 60601 test
Compliance
level
level
Electrostatic
± 6 kV contact
± 6 kV
discharge (ESD)
± 8 kV air
contact
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2kV for power
Not applicable
transient/burst
supply lines
IEC 61000-4-4
± 1kV for input/
Not applicable
output lines
Surge
± 1kV line(s)
Not applicable
IEC 61000-4-5
to line(s)
Not applicable
± 2kV line(s)
to earth
Voltage Dips,
<5% UT (>95% dip
Not applicable
short interruptions
in UT) for 0,5 cycle
and voltage
40% UT (60% dip
Not applicable
variations on
in UT) for 5 cycles
power supply
70% UT (30%
Not applicable
input line
dip in UT) for 25
IEC 61000-4-11
cycles
Not applicable
<5% UT (>95% dip
in UT) for 5 s
Power frequency
3 A/m
3 A/m
(50/60 Hz)
magnetic field IEC
61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Electromagnetic environment-guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the MA801f
requires continued operation during power
mains interruptions, it is recommended
that the MA801f be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should
be at levels characteristics of a typical
location in a typical commercial or hospital
environment.
17