Technical Alarm Description - smartheart 01-554 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ERROR CODES
Code
E0
Pressure system is unstable before
measurement
E1/E2
Failed to detect systolic/diastolic pressure
E3/ E4
Pneumatic system leakage or cuff is too loose/
too tight during inflation
E5
The cuff pressure is above 300mmHg
E6
More Than 3 minutes with cuff pressure above
15 mmHg
E7
EEPROM accessing error
E8
Device Parameter checking error
ER A
Pressure Sensor parameter error

TECHNICAL ALARM DESCRIPTION

The monitor will show 'HI' or 'Lo' as technical alarm on LCD with no delay if
the determined blood pressure (systolic or diastolic) is outside the rated range
specified in part SPECIFICATIONS. In this case, you should consult a physician or
check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory and
cannot be adjusted or inactivated. This alarm condition is assigned as low priority
according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on
LCD will disappear automatically after about 8 seconds.
Meaning
Corrective Action
Re-apply the cuff and attempt new
measurement.
Re-apply the cuff and attempt new
measurement.
Wait 5 minutes. Sit comfortable and remain still;
attempt new measurement. If the error persists,
please contact Customer Care
ENGLISH • 25