CAUTION: Read the manual before use
WARNING: The instrument is not suitable for use in the presence
of explosive or flammable gases, flammable anaesthetic mixtures
or in oxygen rich environments.
CAUTION: Mouthpieces are single patient use. If used on more
than one patient there is a risk of cross-infection. Repeat use may
degrade materials and lead to an incorrect measurement."
CAUTION: Pulmonary filters are single patient use. If used on
more than one patient there is a risk of cross-infection. Repeat use
may increase air resistance and lead to an incorrect measurement."
PLEASE NOTE: The product you have purchased should
not be disposed of as unsorted waste. Please utilise your
local WEEE collection facilities for the disposal of this
product.
PLEASE NOTE: Degree of protection against Ingress of Water is
IPX0
WARNING: To avoid risk of electric shock, this equipment must
only be connected to a supply mains with protective earth
CAUTION: When you connect the SpiroUSB to other equipment,
always make sure that the whole combination complies with the
international safety standard IEC 60601-1 for medical electrical
systems. During measurements, connect the SpiroUSB only to
printers and computers that comply with IEC 60601-1 / ANSI/
AAMI ES60601-1:2005 / CAN/CSA-C22.2 No. 60601-1:14.
Indication for Use
The SpiroUSB spirometer is intended, for prescription use only, to
measure the maximal volume and flow of air that can be moved in
and out of a patient's lungs. The system is intended for use with
paediatric (4 to 17 years) and adult (18 to 99 years) patients in
hospitals, physician offices, laboratories and occupational health
testing environments.
19