Operation
Simply connect the SpiroUSB to a spare USB port on your PC with
the supplied SPCS installed. SPCS will automatically detect that the
SpiroUSB has been connected to a USB port and will run, ready for
you to start your spirometry session. There is no need to remove
the SpiroUSB between sessions.
Please note that with certain PC configurations, even if a
compatible operating system is being used, SPCS may not be
launched automatically when the SpiroUSB is plugged in. In this
case SPCS may be run in the usual way using the Windows on-
screen Start button.
The SpiroUSB transducer is supplied with a two-meter cable. If
this length is insufficient then a remote USB hub should be used.
Remote USB hubs are readily available from computer accessory
suppliers. For some recent computers the USB signal will work
over a longer distance using an extension cable. An extension
cable has been supplied for your convenience, if required. If
communication problems are encountered using the extension
cable, then a remote USB hub must be used.
Note: Keep the PC and Monitor out of reach of the patient at all
times.
Looking after your SpiroUSB Spirometer
Please observe the following precautions:
• Avoid exposing the SpiroUSB to direct sunlight during use.
• Avoid operating the spirometer in dusty conditions or near to
heating appliances or radiators.
• Do not keep the spirometer in a damp place or expose it to
extremes of temperature.
• Do not direct the transducer holder towards a strong light
source whilst operating the spirometer.
21