Cleaning accessories
With the use of a filter MicroGard
IIB (V-892381), MicroGard
IIC
®
®
(V-892384) or a One-way Safety mouthpiece (36-SST1250) for
each patient, cleaning for the components in patient's gas path is
recommended once a month.
When using the disposable cardboard mouthpiece (adult: 36-
PSA1000, pediatric: 36-PSA1200) without a filter and under the
prerequisite that the patient was instructed only to exhale into
the transducer, the following parts have to be cleaned once a day:
adult/ pediatric adapter, transducer.
With any other use as described above all contaminated parts must
be disinfected between patients.
Important note: Used single patient nose clip pads, mouthpieces,
MicroGard
IIB filter and MicroGard
IIC filter must be disposed off
®
®
immediately after the use.
If there are changes on the material surfaces (cracks, brittleness)
the respective parts must be disposed off.