Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
Power
frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The SpiroUSB is intended for use in the electromagnetic environment specified
below. The customer or the user of the SpiroUSB should assure that it is used in
such an environment
Immunity
IEC 60601
Test
test level
Conducted RF
3 Vrms
IEC61000-4-6
150 kHz to
80 MHz
<5% U
T
(>95 % dip in U
)
T
For 0.5 cycle
40% U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5% U
T
(>95 % dip in U
)
T
For 5 s
3 A/m
Compliance
Level
3 Vrms
Not Applicable
Mains power quality
(USB
should be that of a typical
Powered)
commercial or hospital
environment. If the user
of the SpiroUSB requires
continued operation during
power mains interruptions,
it is recommended that the
SpiroUSB be powered from
an uninterruptable power
supply or a battery.
3 A/m
If incorrect operation occurs,
it may be necessary to
position the SpiroUSB further
from sources of power
frequency magnetic fields or
to install magnetic shielding.
The power frequency
magnetic field should be
measured in the intended
installation location to assure
that it is sufficiently low.
Electromagnetic Enviroment-
Guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
SpiroUSB, including any cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
(d)
d = 1.2√P