PEAK MAX/PEAK MIN
Accuracy guarantee range
Range
6.000 V
0.00 V to ±12.00 V
(0.01 V)
60.00 V
±3.0 V to ±120.0 V
(0.1 V)
600.0 V
±30 V to ±1000 V *
(1 V)
1000 V
±50 V to ±1000 V *
(1 V)
*1: Frequency range of 10 Hz ≤ f < 20 Hz is designed value.
*2: Values of up to ±1200 V are displayed, but accuracy is not defined for display values in excess of 1000 V
(which are provided as reference values).
*3: Values of up to ±1700 V are displayed, but accuracy is not defined for display values in excess of 1000 V
(which are provided as reference values).
(Resolution)
2
3
Accuracy guarantee
1
frequency range *
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
Accuracy Table
Measurement accuracy
±1.5% rdg.±0.07 V
±1.0% rdg.±0.07 V
±1.5% rdg.±0.07 V
±1.5% rdg.±0.7 V
±1.0% rdg.±0.7 V
±1.5% rdg.±0.7 V
±1.5% rdg.±7 V
±1.0% rdg.±7 V
±1.5% rdg.±7 V
±1.5% rdg.±7 V
±1.0% rdg.±7 V
±1.5% rdg.±7 V
53
1
2
3
4
5
6
索
引