4. Confirming input
a.
Press the CLEAR button to confirm the input of the adjustment value.
ð
The change from the factory setting is indicated by the CAL symbol, which is now continuously
shown in the display.
If the CLEAR button is not pressed within approx. 15 seconds, the initial state is preserved.
11.1.2 Readjustment
The continuously displayed CAL symbol indicates that an adjustment has already taken place. When a
new adjustment value is entered, this is added to the adjustment value already present.
1. Calculating the adjustment value
The already adjusted instrument shows a new deviation of the mean volume from the nominal vol-
ume (e.g., 0.017 ml). For information on calculating the mean volume, see Checking the Volume (Cali-
bration), p. 66.
2. Entering CAL mode
a.
b.
9974 69
NOTICE
a.
With the instrument powered on, press and hold the
> 3 s
CLEAR button for more than 3 seconds until the follow-
ing modes are repeatedly shown one after another in
the display:
CAL — GLP — APO — dP
b.
When CAL appears in the display, release the CLEAR
button. CAL blinks and the adjustment value of the pre-
vious adjustment appears.