Display System - koban KMOS-01 Guía De Usuario

Multímetro osciloscopio de verdadero valor eficaz
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Oscilloscope Section

2-12 Display System

2 12 1 Time and Date
- -
•After startup, actual time and date display at the top
left corner on the interface, format: mm/dd/yy,
hh/mm/ss (picture 2-10-1).
•Function of the clock is supplied by the back-up
battery inside of the device, which can work for 5 to
10 years, and irrelevant to the Li-ion rechargeable
battery.
•Take "Operating Instructions For Oscilloscope" for
reference for the adjustment of clock.
2 12 2 Battery Electric Power Display
- -
•After startup, at the top right corner, display the working
Li-ion rechargeable battery electric power (picture 2-10-2).
•The battery fully charged, 4 cells.
•When 1 cell left, blue color turn to red, warning to
recharge, 30 to 45 minutes may be used by this left
power.
•All cells disappear, please recharge promptly, 15 to
20 minutes may be used by this left power.
2 13 Measuring System
-
The oscilloscope displays forms of voltage relative to time, and helps user measure and display
waveforms. Two ways of measurements can be adopted: Scale measurement or cursor
measurement.
2 13 1 Scale Measurement
- -
Fast and direct evaluation can be done by this way, for example, waveform amplitude can be
observed to judge whether it's 0.8V or not.
By calculating relevant primary and secondary scale division, then multiplying by proportion
factor, simple measurement can be taken.
For example, if four main vertical scale divisions between the waveform peak and base are
figured out, proportion factor is 200 mV/division, peak value voltage can be calculated by the
following method: 4 divisions * 200mV/div= 0.8V
2 13 2 Movement of Cursor
- -
As the following picture shows, press "HOLD" or "REC", "V/Fcursor" appear in prompt bar,
indicates function switch button for cursor measurement.
1. Blue vertical cursor appears on the displayed screen, press "
time for one space, press this key for more than 1 second for fast move, the cursor moves
continuously in the same direction.
2. Press "MODE" to switch left/right cursor, lighter cursor indicates state of activation.
3. Display mode of frequency and cycle is also switched at the time of left/right cursor switch.
Activate left cursor displays frequency, right cursor for cycle. See picture 2-11-1.
48
Picture 2-10-1
Picture 2-10-2
" to move the cursor, one

Publicidad

Tabla de contenido
loading

Tabla de contenido