Aesculap Power Systems
Acculan® 3Ti Dermatome GA670
23
20
Fig. 4
Installing the sterile funnel in the battery
compartment
Have a helper insert the (unsterile) battery 23
through sterile funnel 20 into the battery
compartment (observe color code 26).
Note
Once the battery is in the battery compartment, a series
of acoustic blips will signal that the dermatome is ready
for operation.
Note
The color code on the floor of the battery compartment
must match the color code at the bottom of the battery.
8
26
Fig. 5
Removing the sterile funnel
Once the battery has been inserted, have a helper
remove sterile funnel 20 (unsterile), see Fig. 5.
20