Slightly shake dermatome 1 with battery removal
device 25 mounted and the battery compartment
facing downward.
Battery 23 smoothly slips into battery removal
device 25.
Hand battery removal device 25 containing the
discharged battery 23 to an unsterile helper.
Insert charged battery 23, see Inserting the
battery.

Protection against inadvertent activation

4
5
Fig. 10 Engaging the safety catch
To prevent inadvertent activation of dermatome 1
while replacing the dermatome blade, trigger 4 must
be locked.
To lock trigger 4: Move safety catch 5 to its OFF
position, see Fig. 10.
Trigger 4 is locked. Dermatome 1 cannot be activated.
To unlock trigger 4: Move safety catch 5 to its ON
position, see Fig. 10.
Trigger 4 is unlocked. Dermatome 1 can be
activated.
1

Intraoperative storage

Fig. 11 Storing the dermatome ready for operation in
its Eccos® holder
Note
Eccos® holder can be used for intraoperative storage of
the dermatome. The dermatome cannot be processed in
this position. For the correct processing position, see
Mechanical cleaning/disinfecting!
Engage safety catch 5 against inadvertent
activation, see Protection against inadvertent
activation.
Fold back the clamp of the Eccos® holder.
Place dermatome 1 in the Eccos® holder.
11