Aesculap Power Systems
Acculan® 3Ti Dermatome GA670
5.7 Mechanical cleaning/disinfecting
Note
Categorically, the disinfector must be of tested and
approved effectiveness (e.g. DGHM or FDA approval or
CE mark according to DIN EN ISO 15883).
Note
For thermal disinfection, always use fully desalinated
(demineralized) water. Ensure that Ao is >3 000 for the
process.
Note
The disinfector used for processing must be serviced
and checked at regular intervals.
Fig. 13 Tray with product in Eccos® holder
Dismantle the flap rod, see Dismantling the flap rod
of the Acculan® 3Ti Dermatome.
Store flaps and nut separately.
Ensure that the blade cover is kept open through
the entire processing procedure.
Use Eccos® tray GB256R fitted with Acculan® 3Ti
holders, or mount appropriate holders (e.g. Eccos®
holder GB498R) in a suitable tray (instructions for
use TA009721 for Aesculap Eccos® storage
system).
18
Insert the product in its proper position, with the
trigger pointing upwards, in Eccos® holder
GB498R.
Place the battery compartment over the clamp and
swivel the device into the Eccos® holder so that the
trigger points upward, see Fig. 13.