3.5

Installing the ceiling arm

WARNING
Injury hazard from spring balanced arm jumping
Do not remove the transport safeguard until after the
luminaire head has been installed.
B
A
C
Carefully turn the covers "A" and remove them from the
joint cover "B".
Loosen the screws "C" and remove the joint covers "B"
from the transverse boom.
D
Carefully remove the rotary plug "D".
Insert the ceiling arm into the ceiling tube.
A
After inserting the ceiling arm, first install the ring "A",
then the locking ring "B", followed by the end rings "C".
D
Carefully insert the rotary plug "D".
B
A
C
Attach the joint covers "B" and install the screws "C".
Insert the covers "A" and turn to secure.