INSTALLATION: Triango 80 luminaire
head
DANGER
Death hazard from electric shock.
►
Before any work on the device, disconnect the current
supply or pull the plug out of the socket and secure
the device against being turned back on.
WARNING
Injury hazard
►
The spring balanced arm is under high spring stress.
If the boom is lowered without the installed device, it
MUST be held firmly.
►
If it is let go, it springs upward and can cause severe
injury.
►
Never remove the luminaire body unless the boom is
in the upper position or is held securely in low position
by a second person.
►
To prevent severe injury or damage, always get the
help of a second person when installing and removing
the luminaire body.
Remove screw "A" and pull off the yellow protection cap
►
"B" from the spring arm safety device.
►
Attach the luminaire head.
►
Turn the sleeve to insert the locking device.
34
A
B
►
Turn the sleeve back the other way and secure with a
screw.
WARNING
Injury hazard from falling luminaire head.
►
Make sure the locking device is properly installed.
►
Damage to due to a poorly mounted luminaire head.
Tighten the screw as the luminaire head's friction
►
requires.
►
Attach the hand grip.
7.
OPERATION
Triango 80 C, W, F
CAUTION
This product emits potentially hazardous optical
radiation. Do not stare directly into the light cone. Eye
irritation may occur.
The radiation emitted by this product complies with the
exposure limits for reducing the risk of photobiological
hazards based on IEC 62471: RG 2 (medium risk).
►
Turn the luminaire on with the rotary knob.
►
The luminaire can be dimmed by turning the head.
►
Before each use, perform a functional test: All LEDs in
the light cone must come on.
CAUTION
►
The rotary knob is not removable.