Installing The Wall Mounting Arm - Derungs Dmed Triango 80 C Instrucciones De Montaje Y Funcionamiento

Luz de tratamiento
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 80
D
Insert wall bracket "A" (with end ring "C" and cover "B" in
position) into the wall holder "D" and simultaneously pull
the plug through the rectangular cutout.
E
F
When the wall bracket is vertically aligned, install it with
the M8 locking screw and nut "E" and tighten (20 Nm).
Tighten all 4 grub screws "F" (5 Nm).
Connect the ceiling tube plug with the power supply plug.
Push the cover against the wall and screw tight with the
ring (0.5 Nm)
C
B
A
4.3

Installing the wall mounting arm

WARNING
Injury hazard from spring balanced arm jumping
Do not remove the transport safeguard until after the
luminaire head has been installed.
B
A
C
Carefully loosen the cover "A".
Loosen the screws "C" and remove the joint covers "B".
D
Carefully remove the rotary plug "D".
Install the wall mounting arm.