Tighten the cable holder using a type 3 Allen screw on
the upper support tube (2.4 Nm).
WARNING
Never unscrew the upper cable holder — injury hazard.
If both cable holders are unscrewed, the connecting
piece will come loose and fall down, which can cause
personal injury and damage to the cable and device.
Carefully pull off the covers "A".
Remove the M4 Allen screw with the washer.
Push the spring balanced arm's plug against the upper
support tube's plug until they click together.
A
Attach the spring balanced arm to the support tube.
Align the thread in the spring balanced arm according to
the opening in the support tube, insert the M4 Allen
screw and washer, and then tighten.
A
Reattach the covers "A" one after the other. Make sure
that the cover "B" is positioned in the groove of both
covers "A".
DANGER
Death hazard from electric shock.
Do not turn on the power supply before the luminaire
head has been installed.
ENG
B
33