Triango 80 F
DANGER
Death hazard from electric shock.
►
Do not plug in any damaged power cables.
►
If the power cable shows damage, immediately replace
it with a new one.
►
The connection voltage and frequency must match
the data on the type plate.
►
Only connect to power grid with protective ground
conductor (PE)
CAUTION
When changing locations
►
Put the terminal into its lowest position.
►
Unlock the casters.
►
Take care not to run over anything, including the
connection cable.
►
Hold firmly during transport within the clinic.
►
Be particularly cautious of slopes, thresholds, bumps
and other obstacles.
CAUTION
►
When it is not in use, wind the power cable onto the
cable holder.
►
Insert the power cable.
►
Connect cable to the power grid.
8.
CLEANING AND DISINFECTION
DANGER
Death hazard from electric shock
►
Before disinfection cleaning, switch off the power
connection and secure against unintentional activation.
CAUTION
Material damage due to incorrect cleaning
►
For cleaning, use only products that do not impair
operation of the luminaire.
►
For cleaning, do not use solvent-based, chlorine-
based or abrasive detergents, because they can crack
plastic parts and cause other damage.
►
The cleaning agents must be approved for use on
plastics such as PC, PMMA, PA and ABS.
►
Concentrated disinfectant can damage the luminaire.
►
For concentration and application times, check the
information provided with the product used.
►
The wrong cloth may cause scratches.
RECOMMENDED DISINFECTANTS
►
Dismozon Plus
►
Kohrsolin Extra
►
Lysoformin
►
Microbac Tissues
►
Mikrozid Sensitive Liquid
CAUTION
Dirt reduces the light strength
►
Keep cover clear through regular cleaning.
►
Only wipe cleaning allowed.
►
Clean the PMMA clear cover with a chamois soaked in
windshield wiper fluid.
CAUTION
To minimize the risk of disease transmission, applicable
health and safety regulations and the requirements of the
national hygiene and disinfection authorities must be
observed in addition to these instructions.
ENG
35