Step by Step Instructions
6.5.7
Apical Reverse Functionality
Apical Reverse function allows the X-Smart IQ
opposite direction, when the relevant position for working length determination is reached.
Apical Reverse function is not activated by default in the X-Smart IQ
function is only available when the X-Smart IQ
App and only works in combined mode with Propex IQ
Activation or deactivation of Apical Reverse function is described in detail in the
®
Endo IQ
See Endo IQ
6.5.8
Stand-by Mode
X-Smart IQ
time.
Mode
Device active
Stand-by mode 1
(inactive for 5 min)
Stand-by mode 2
(inactive for 35 min)
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
App User Manual.
®
App User Manual -
®
Handpiece enters stand-by mode when it is not used for a certain period of
LED description
All LEDs are active.
®
X-Smart IQ
Handpiece START/
STOP button flashing
blue every 3 s.
All other LEDs are
OFF.
All LEDs OFF.
®
Handpiece to automatically rotate in the
®
Handpiece is connected to Endo IQ
®
6 Step by Step
Instructions.
®
Endo IQ
App
display (if
applicable)
®
Bluetooth
ON, device
visible on status bar.
®
Bluetooth
ON, device
visible on status bar.
®
Bluetooth
OFF,
device not visible on
status bar.
®
Handpiece. The
Apex Locator.
Wake-up action
None.
®
(X-Smart IQ
Handpiece ready for
use).
Pick-up X-Smart IQ
Handpiece.
Press the multifunction
button.
®
®
21/52