Technical Characteristics
Power supply
Charger power supply
Frequency
Charger nominal power input
Torque range
Speed range of the micromotor shaft
Type of protection against electrical shock
Applied part
Safety level in the presence of flammable anesthetic
gas mixtures or oxygen
Mode of operation
Ambient conditions
Medical device class
Conditions for transport and storage
34/52
Specification
Lithium ion battery:
3.6 V, 2,000 mAh
100 - 240 VAC
50 - 60 Hz
5.5 VA
5 mNm – 50 mNm
(0.5 Ncm – 5 Ncm)
Dependent on motor speed
1 200 – 5 100 rpm (in rotary mode)
Independent of contra-angle reduction
Class II and internally powered
equipment
BF (contra-angle)
Not suitable for use in the presence of
flammable anesthetic gas mixtures or
oxygen
Non Continuous Operation
Duty Cycle: ON (Maximum activation
time): 3 mins; OFF (minimum
deactivation time): 1 min
Use: in enclosed spaces
Ambient temperature:
15°C - 35°C
(59°F - 95°F)
Relative humidity:
< 80%; non-condensing at 0°C
(32°F)
Operating altitude:
< 2 000 m (approximately 6561 ft)
above sea level
Maximum housing temperature
according to IEC 60601-1:2006
Class IIa, Rule 9, according to MDD
93/42/EEC (Annex IX)
Ambient temperature:
-20°C - +50°C
(-4°F - 122°F)
Relative humidity:
20% - 80%, non-condensing at
> 40°C (104°F)
Atmospheric pressure:
50 kPa – 106 kPa
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01