Hand Piece Cartridge Heater - If gutta-percha gets into the cartridge
heater section of the hand piece, turn on the unit and press the Return
button on the console to fully retract the plunger. Allow the heater section
to reach operation temperature (180°C). Turn off unit. Insert the included
hand piece cleaning brush into the heating chamber. Rotate the brush
several times to remove the gutta-percha from the chamber.
Hand Piece Delivery Plunger - Annually: Without a cartridge inserted,
press the activation cuff until the gutta-percha indicator moves all the
way forward. Then, push the Return button and allow the indicator to
completely retract.
Hand Piece Cartridge Nut - Allow the Flow hand piece cartridge nut to
cool before removing any excess gutta-percha. The cartridge nut may be
steam autoclaved for 18 min at 134°C.
Cartridges - Cartridges are intended for single patient use. Prior to using
the device on a patient, wipe the cannula with alcohol or a disinfectant.
(Make sure that the device has been switched off and cooled down.) Store
cartridges at room temperature. Do not immerse the cartridges into any
liquid. Dispose of cartridge in a biohazard container. The cartridges must
not be used after the expiration date!
Heat Shield - Steam autoclave for 18 minutes at 134°C.
Note
All liability is disclaimed for failure to follow these instructions or use of
non-validated methods for the re-use of instruments!