5 Operation
5.3.
26
NightKNIFE / Ligator
Changing the output power of the HF device
Before increasing the output power of the HF device,
check that:
all HF cables and connectors are properly
connected;
the ligation instrument is properly connected
(see the operating instructions of the HF
device);
the foot switch works properly;
the insulation of the HF cable D, the jaws
the insulation of the HF cable C3, the jaws
A1, the shank tube B and the trocar sleeve
is in good condition;
the active electrodes of the jaws A1 are
clean and not worn out.
Withdrawal
WARNING
Risk of patient injury due to damaged or broken-off
parts!
Check the jaw insert after each withdrawal. All parts
must be present.
1.
Close the jaws A1.
2.
Withdraw the ligation instrument from the trocar sleeve.
MN031-447-S1 EN
MN031-447-S0-20101217-EN