Step
Picture or additional info if applicable
Step 4 – Connect Connector Cable
Disconnect the Charger Module from the CHARGER CONNECTOR if it is
connected.
Slide the CONNECTOR DOOR fully to the left, revealing the THERAPY
CONNECTOR as shown.
Connect the GENERATOR CONNECTOR end of the Connector Cable to
the THERAPY CONNECTOR. Orient the connector and gently push in.
Slide CONNECTOR DOOR to left to
The connector is magnetic and will engage as the magnet gets close.
reveal the THERAPY CONNECTOR
Push gently to confirm connector is fully engaged.
THERAPY CONNECTOR
Step 5 – Prepare Lithoplasty Catheter for Use
N/A
Prepare the catheter for use following the instructions in the
Lithoplasty Catheter IFU. Use a sterile sleeve to cover the distal end of
the Connector Cable. Protect the connector from contamination by
fluids.
Step 6 – Connect Lithoplasty Catheter
Take care not to contaminate either connector end with fluids or other
foreign matter during this procedure prior to mating.
Connect the Lithoplasty Catheter to the CATHETER CONNECTOR end of
the Connector Cable using the same method outlined in step 4.
NOTE: Ensure that the sterile sleeve also covers the CATHETER
CONNECTOR.
The THERAPY STATUS indicator on the Generator front panel and the
CATHETER CONNECTOR will be yellow, indicating that the Generator is
ready to activate.
The catheter balloon dimensions will appear in the BALLOON SIZE
display.
Step 7 – Position the Lithoplasty Catheter
(Refer to the Lithoplasty Catheter IFU)
Following conventional angioplasty catheter techniques, introduce and
position the Lithoplasty catheter as desired. Use caution to prevent
unintentional movement of the Connector Cable and Lithoplasty
Catheter during treatment.
LBL 60780 rev. B