Step
Picture or additional info if applicable
Step 5 – Check Diagnostics
Normal
When powered on, the Generator will perform an array of built in tests
designed to detect certain malfunctions. If an error is detected, an error
message will be displayed. If there are no error messages, these tests
passed successfully.
If an error message is displayed, refer to Troubleshooting, Section 7.0.
Error Condition
Step 6 – Initiate the Output Test
This test is initiated manually by depressing and holding the THERAPY
ON/OFF button and releasing this button when the THERAPY STATUS
indicator comes on green. Pressing this button for three seconds is
required.
Step 7 – Confirm Output Test Result
The output test requires approximately 15 seconds to complete. During
this time the THERAPY STATUS indicator will remain on green. Upon
successful completion of this test, the Generator will sound four beeps. If
an error is detected, an error message will be displayed. If the display
remains blank with only the battery symbol, this test has passed
successfully.
This is the final step of the recommended inspection and test procedure.
LBL 60780 rev. B