Testing The Generator - Shockwave Medical Lithoplasty 825D Manual Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 227
WARNING
POSSIBLE DEVICE SHUTDOWN
This device operates only from an internal battery source. Charge the Generator battery when not in use. Always operate the
Generator when the battery symbol is at least ¼ Full. Generator may shut down without warning when the battery symbol is
Empty. Remove the Generator from use and contact your Shockwave Medical representative in the event the battery symbol
frequently appears Empty or if the battery symbol does not appear full after twelve hours of charging.
6.1.2

Testing the Generator

The Generator will automatically perform an array of built in tests designed to detect certain malfunctions
each time it is turned on. In addition, the Generator features an automated test of the Lithoplasty output
system which may be initiated by the user. Confirmation of satisfactory test results is recommended daily,
before cases are initiated, or as directed by your Biomedical Department. These tests may be done as follows:
Step
Step 1 – Turn the Generator On
Momentarily depress the POWER ON/OFF button. All indicators on the
Generator will come on briefly as a test. The THERAPY STATUS indicator
will come on green and yellow during this test. The POWER STATUS
indictor will continue to be on green.
Step 2 – Check Diagnostics
When powered on, the Generator will perform an array of built in tests
designed to detect certain malfunctions. If an error is detected, an error
message will be displayed. If no error is displayed, these tests passed
successfully.
If an error message is displayed, refer to Troubleshooting, Section 7.0.
Step 3 – Initiate the Output Test
The Charger Module must be connected to AC Mains and to the
Generator CHARGER CONNECTOR to run this test.
This test is initiated manually by depressing and holding the THERAPY
ON/OFF button and releasing this button when the THERAPY STATUS
indicator comes on green. Pressing this button for three seconds is
required.
LBL 60780 rev. B
Picture or additional info if applicable
24 |
EN
P a g e

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

M732lg825d1

Tabla de contenido