22
3.
Setting up the unit
3.1. Installing the battery
OSIRIS units are supplied with the battery not installed.
To install the battery:
• Remove the cover plate 22 ,
• Connect the battery and insert it in its housing,
• Reinstall the cover plate, ensuring that it is fully engaged.
3.2. Connections
Oxygen supply to OSIRIS 3
• Connect the OSIRIS 3 input to a source of oxygen via the
O
supply tube.
2
If the oxygen is from a cylinder, the cylinder must be
equipped with a reducer adapted to supplying medical
oxygen.
The nominal supply of O
bar) and 600 kPa (6 bar).
Connect the gas supply hose to the ventilator connector
before connecting it on the network (or cylinder) side.
To disconnect (after shutting down the unit), start by
disconnecting the hose from the network (or cylinder)
connector.
Electrical supply connection
• When running off the mains supply, connect the special
electrical unit (supplied by Air Liquide Medical Systems).
The electrical powersupply also charges the internal
battery, whether OSIRIS 3 is in use or not (battery charging
time: 3 hours).
Setting up the patient circuit
Two types of patient circuit can be used on the OSIRIS range.
Single-use patient circuit
A single-use patient circuit, composed of:
• a single-use patient tube (1) together with its control tube
(2),
•
a single-use expiratory valve already connected to
the tubes above (3).
All our single use patient circuits are assembled in their
original packaging.
Setting up
• Connect one end of the ringed tube onto the outlet of the
OSIRIS 3 (standard conical outlet, 22 mm diameter).
• Whenever possible, fit a new bacteriological filter onto the
"patient" output of the expiratory valve.
YL033700 - Rev. 3a - 2019-02
Setting up the unit - EN
must be between 280 kPa (2.8
2
11