4.4. Cleaning, Disinfection, Sterilization
4.4.1. Single-use patient circuit
A single-use patient circuit, which is composed of:
• A single-use patient tube (1) together with its control tube
(2);
• A single-use expiratory valve already connected to the
tubes above (3).
All our single use patient circuits are assembled in their
original packaging.
Three single-use patient circuits are delivered together
with the ventilators.
4.4.2. Autoclavable patient circuit (optional)
• Clean the patient circuit with soapy water or with a clean-
ing solution..
Do not use powerful solvents (ether, acetone, etc.).
• Disinfection of the patient circuit may be made by immer-
sion in a disinfecting solution, after cleaning.
• Steam sterilize the patient circuit, following the instruc-
tions in the adjacent table.
• Bacteriological filter: It is advisable to change the
• bacteriological filter on the insufflation output connection
at least each time the patient circuit is sterilized, or each
time the equipment is connected to a new patient.
4.4.3. OSIRIS 3
• The unit may be cleaned with a cloth dampened with
soapy water, avoiding any risk of water entering the appara-
tus (electronic circuitry may be damaged), then wipe with a
dry cloth or a cloth impregnated with a water/ alcohol solu-
tion.
• Disinfection of the outer surfaces of the unit may be
made with a spray such as Surfanios 
the manufacturer's instructions, especially dosage) or with
cloths impregnated with an alcohol-based disinfecting solu-
tion.
• OSIRIS 3 may be sterilized with a formaldehyde gas unit.
Proceed as follows:
• Connect the OSIRIS 3 to the sterilizing unit via the ambiant
air connection
on the rear of the ventilator,
• Set button f to 20 c/min,
• Run the sterilizing unit according to its manufacturer's in-
structions.
It is imperative to use air which is extremely clean and free
of any trace of oil or water.
Sterilization with formol using any other procedure than
that described above may cause formol crystallization and
stop OSIRIS 3 working.