Set-up and Start-up
Transportation
safety device
14
4.2 Transport
The motor lift units have a transportation safety device, which must be removed prior to
the initial start-up of the unit or reattached prior to the return shipment. This is very im-
portant in shipping any return or repair of motor lift unit to make sure transport lock is
assembled.
A brief description of the procedure for removing and reattaching the transportation safe-
ty device may be found in the following.
The transportation safety device consists of three M5x8 screws and a connecting plate
with holes for fixing the position.
1
2
Figure 4-1: Transportation safety device
1
Transportation safety device
2
Front shell of lift
3
Power switch
4.2.1
Removing the transportation safety device
All hose and cable connections to other devices, and the coolant/vacuum connec-
tion are disconnected.
Connect the unit and set the power switch (3) at on.
1.
Remove the top screw (in the front shell of the lift (2)) and the lower screw of the
transportation safety device (1) in the switched state. Never tighten in the top posi-
tion!
2.
Remove the screws and the transportation safety device (1) and store.
3.
Move the hoist up by using the "lift up" (5) button. Switch the unit off and on.
4.2.2
Attaching the transportation protection device before packing
All hose and cable connections to other devices and the coolant/vacuum connection
are disconnected.
Connecting the unit
1.
Continue pressing the lift position contact down when switching the unit on (this will
suppress the "lift to the top when power off" function).
2.
Move the hoist down by using the "hoist down" (4) button
Please pay attention to the lift-height stop!
3.
Hold the transport safety device as shown in Image 4-1 and fix it with the screws
M5x8 (3 pieces).
3
4 & 5
4
Button - lift down
5
Button - lift up