WARNING
Imploding glass devices!
Serious injuries from shattering glass.
Check glass devices for damages (stars, breaks, etc.).
Only use perfect glass devices and replace all damaged
glassware as soon as possible.
6.7 SET
mode
pressure
The vacuum is controlled and maintained at the set pressure in the SET
6.7.1
Menu display
The following values can be changed in the menu view:
Heating bath
Setting of the bath temperature. Setting range is between 20 °C and
maximum heating bath temperature (MAX TEMP is adjustable in
Setup) with a highest range of 210 °C.
Rotation
Setting of rotation speed of the distillation flask. Setting range is be-
tween 10 and 280 rpm.
Vacuum
Setting of the absolute pressure in the rotary evaporating system.
Setting range is between 1 and safety pressure (SAFETY PRES-
SURE is adjustable in Setup) with a high limit of 1200 mbar.
Max. power
Vacuum pump (Vario Models) throughput % to maximize pump
pump
power. Setting range is between 10 and 100 %. Reduction of the
throughput leads to higher solvent recovery rates.
Hysteresis
The value of the hysteresis indicates how much the pressure may
rise (in mbar) before the vacuum valve is opened. Setting range is
between 1 and 50 mbar (only by using valve controlled pumps Rota-
vac valve control, Rotavac valve tec, or House vacuum source).
Timer
Setting of the timer to shut down the process. Setting range is be-
tween 1 and 1440 minutes. For continuous operation choose \"off\"
for Timer setting. The heating bath at the end of a timed run can be
set to turn off (Parameter found in Setup).
Selecting the entry by turning the Hei-GUIDE (6)
1.
Call up the selected entry to be changed by pressing the Hei-GUIDE (6) (entry will
be highlighted in orange).
2.
Set the value by using the rotary pushbutton controller Hei-GUIDE (6).
3.
Pressing it again confirms the change. Softkey "Esc" (1) terminates the change of
the currently processed entry.
4.
The new settings are accepted by the "accept" entry. It is changed to the actual
value display.
All changes in the menu view are discarded with softkey "Esc" (1).
The parameters are stored under a name to be selected by using softkey "Store
Favorite" (1).
Hei-VAP Precision
Operation of the Hei-VAP Precision
mode.
pressure
EN
49