EN
Operation of the Hei-VAP Precision
60
6.18 Setup
Online Help
Shows Online-Help instead of the actual values in the menu
SETpressure, AUTOeasy, AUTOaccurate, Gradient, Favo-
rites and Setup.
Safety pressure
If the system reaches the setting of the Safety Pres-
sure point, the vacuum control and flask rotation is
stopped, the evaporator is vented and the motor lift is
raised (ML only). Setting range is between 900 and
1200 mbar.
Heating Bath Media
Allows switching between different heating bath media
and sets a limit to the maximum heating bath tempera-
ture for the media for operator safety.
Max. Temp. Heating
Sets the limit of the maximum heating bath tempera-
Bath
ture which can be set. The chosen bath media has an
influence on the temperature range. Maximum set
range between 20 and 210 °C.
Language
Sets language of the interface.
Bath Power at End
Sets the state of the bath after the process has ended.
When setting is \"off\" the bath is powered down after
the process is terminated.
Brightness Display
Changes brightness of the display.
Calibrate p-sensor
Opens sub-menu for calibration of the pressure sensor.
Error history
Shows the errors and history of occurrence for service data.
Reboot setup
All settings of the Setup are put into the state of factory
setting when Reboot enganged.
Scale time axis
Sets range of the visible time frame in the graphical
\"display actual values\". Set range is between 1 and
120 minutes.
6.19 Error messages
Error messages are only displayed in the event of an error. The error text and its trouble-
shooting are explained in chapter 7, "Errors and Troubleshooting".
Hei-VAP Precision