9. Bluetooth
configuration
®
E
The connection procedure has been designed to be simple and is signalled by the following three states :
1° Symbol
off .............. disconnected mode
2° Symbol
blinking ...... advertising mode
3° Symbol
on .............. connected mode
Connexion :
1° Activate Bluetooth ® Smart compatible software and hardware (Master : PC, Display Unit)
2° Start the instrument. By default the Bluetooth® module is active and the instrument is available for connection during
120s (advertising mode)
3° As soon as the device is detected, a connection is established automatically. If no connnection is established during
120s, reactivate the Bluetooth ® module using the bt / On menu.
4° Instrument is ready to communicate (connected mode)
Pairing :
Pairing with master is automatically done at first connection.
To connect the instrument to a new master (new pairing), it is necessary to clear all pairing information on the instrument
using the bt / rESEt menu.
14