11. Use with calibration to the MIN (or MAX) measured value
Certain applications need to calibrate the instrument to the MIN (or MAX) measured value. In this case, proceed as follows :
11.1 Calibration of the instrument
- Enter a Preset value corresponding to the actual size of the standard (see chap. 3)
- Select the MIN mode (or MAX, depending on the application) (see chap. 4)
- Make a standard measurement (going through the turnaround point)
- Calibrate the instrument by selecting the SEt mode and pressing on the button
- The instrument is calibrated and ready to measure.
11.2 Measurement
- Make the measurements. The digital display stores and displays the MIN (or MAX) measured value.
- Before each new measurement, reset the measured value by a short press on
17
(see chap. 3)
SET
E