OBJ_DOKU-197851-002.fm Seite 28 Dienstag, 13. April 2021 2:46 14
5.6

Check

1. Check the product for visible damage and wear:
– Silicone surface intact, no tears, holes, kinks etc.
– Connector not broken or corroded
– Discolorations on the silicone surface are harmless.
If damaged, do not use this product!
5.7

Overview of validated procedures

The following procedures have been validated as suitable for
processing the product:
• Cleaning by machine and disinfection in a Uniclean PL-II 15-
2-EL disinfector (MMM GmbH, Munich, Germany) with the
cleaning agent neodisher mediclean forte (Dr. Weigert
GmbH & Co. KG, Hamburg, Germany).
– neodisher mediclean forte: Dosage 0.5%, 5 minutes ex-
posure time at 55 °C
– Disinfection: 5 minutes at 90 °C.
6

Monitoring of the neutral electrode by the

electrosurgical unit

Erbe VIO electrosurgical units are equipped with a Neutral Elec-
trode Safety System (NESSY), which monitors the neutral elec-
trode, warns of critical situations, and thus prevents burns. Ob-
serve the unit's optical and acoustic warning signals. Observe the
error and advisory messages from neutral electrode monitoring.
With this neutral electrode type, the electrosurgical unit moni-
tors three safety-related parameters:
• the connection with the electrosurgical unit
• the contact to the patient's skin
• the application direction of the neutral electrode (NESSY
symmetry monitoring)
28