Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
EN
WARNING
Only use the input voltage and frequency as rated on the product.
Always allow the product to reach room temperature before you begin setup or test functional operations, to
prevent permanent product damage.
Do not use this product if it has identifiable failures, defects, malfunctions, or damage.
Do not use this product under any conditions if use would cause injury to the operator or patient.
Only operate the product when all operators are clear of the mechanisms.
Do not attach the power cord to any parts of the product.
Always unplug the power cord and call maintenance if unanticipated motion occurs.
Do not store items under the bed.
Do not use the bed without the litter covers.
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
Always store the power cord to avoid the risk of entanglement, damage to the power cord, or potential shock
hazards. If the power cord is damaged, remove the product from service and contact the appropriate maintenance
personnel.
Always allow enough clearance between the head end of the product and the adjacent wall, so you can unplug the
power cord from the wall outlet during an emergency.
Always store the power cord before transporting the product.
Always disconnect the power cord from the wall outlet if you detect overheating of the battery, control cables, or
pendants. Do not use the product again until it has been inspected, serviced, and confirmed to work as intended by
authorized maintenance personnel.
Always replace the battery after it surpasses its expected service life.
Do not open a dead battery.
Do not throw the battery into a fire.
Do not spill liquid onto the battery or submerge the battery in liquid.
Always unplug the battery cable from the control box before you store the product for a long period of time.
Always lock the siderails in the full up position with the sleep surface horizontal in the lowest position when
transporting a patient.
Always keep limbs, hands, fingers, and other body parts clear of mechanisms and gaps.
Always make sure that there are no obstacles near the product. Injury to the patient, operator, bystanders or
damage to the frame or surrounding equipment could occur if you collide with an obstacle.
Do not attempt to move the product laterally. This may cause the product to tip.
Do not move the product after you apply the brakes.
Always apply the brakes when a patient is getting into or out of the product to avoid instability.
Always apply the brakes when the patient is unattended.
Always make sure that all persons and equipment are away from the area below and around the backrest before
you activate the CPR release. The CPR release is for emergency use only.
Always properly orient the headboard when replacing the headboard to avoid entrapment.
Always properly orient the footboard when replacing the footboard to avoid entrapment.
Always make sure that all persons and equipment are away from the area below and around the leg rest before you
lower the lower leg section.
Always lock the siderails unless a patient's condition requires extra safety measures.
Always lock the siderails in the full up position when the patient is unattended.
1-4

Summary of safety precautions

KK-7500 REV 02
www.stryker.com