WARNING (CONTINUED)
Do not use siderails as restraint devices to keep the patient from exiting the product. The operator must determine
the degree of restraint necessary to make sure that the patient is safe.
Do not sit on the siderails.
Always lock bed motion controls when the patient is unattended.
Never store the nurse control panel within the patient's reach.
Do not sit on the bed extender. This may cause the product to tip.
Always lock the bed extender before you place weight on the bed extender.
Always retract the linen tray (option) before you put the product in motion.
Always retract the linen tray (option) when not in use.
Do not use the product for X-ray procedures if it does not have the radiolucent backrest (option).
Do not use accessories to support patient limbs or other body parts.
Do not clean, service, or perform maintenance while the product is in use.
Always power off and unplug the power cord before cleaning, servicing, or performing maintenance.
Always power off the product and unplug the power cord from the wall outlet when large spills occur near the circuit
boards, cables, and motors. Remove the patient from the product, clean up the fluid, and have service personnel
inspect the product. Fluids can cause unpredictable operation and decreased functionality of any electrical product.
Do not return the product to service until it is completely dry and has been thoroughly tested for safe operation.
Do not spray cleaners directly onto the battery, control boxes, actuators, cables, or other electric equipment.
Do not use abrasive powder, steel wool, or similar materials that may damage the product surface.
Do not use Virex
®
TB for product disinfecting.
Do not use acid-based chemicals or flammable chemicals, such as gasoline, diesel, or acetone for cleaning
purposes.
The cleaners and disinfectants must not be highly alkaline or acidic (pH value 6-8).
Do not directly spray or saturate the siderail control panel, patient control pendant, or nurse control pendant with
cleaners.
Do not use sharp objects to clean the siderail control panel.
CAUTION
Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in
this manual.
Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to patient or operator. Modifying the product also voids its warranty.
To minimize the risk of any electromagnetic interference, the product design follows the standard IEC 60601-1-2.
To avoid problems, use the bed in accordance with the EMC/EMI requirements in the EMC section of this
operations manual.
Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 in.) to any part of the SV2, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
www.stryker.com
Summary of safety precautions
KK-7500 REV 02
English