Guidance and manufacturer's declaration - electromagnetic immunity

The [MDS3003]/[MDS3003P] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [MDS3003]/[MDS3003P] should assure that it is used in such an environment.
IMMUNITY
IEC 60601test level
test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power
Frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Compliance level
± 6 kV contact
± 8 kV air
3 A/m
-28-
environment - guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Power frequency magnetic
fields should be at levels
3 A/m
characteristic of a typical
location in a typical commercial
or hospital environment.