Medline MDS3003 Guía De Operación página 30

Ocultar thumbs Ver también para MDS3003:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Radiated RF
3 V/m 80 MHz
IEC 61000-
to 2.5 GHz
4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the [MDS3003]/[MDS3003P] is used exceeds the applicable RF compliance level above, the [MDS3003]/
[MDS3003P] should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the [MDS3003]/[MDS3003P].
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF
3 V/m
transmitters, as determined by an electromagnetic site
survey,
should be less than the compliance level in each
a
frequency range.
equipment marked with the following symbol:
-30-
Interference may occur in the vicinity of
b

Publicidad