magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test
T
level.
Guidance and manufacture's declaration – electromagnetic
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration – electromagnetic immunity
The CMS70A is intended for use in the electromagnetic environment
specified below. The customer or the user of CMS70A should assure
that it is used in such an environment.
Immunity
test
Conducted
RF
IEC 61000-
4-6
Radiated
RF
IEC 61000-
4-3
immunity –
IEC
60601
Compliance
test
level
level
3Vrms
3 V
150
(for main
kHz to
power line)
80
MHz
1V
outside
(for probe)
ISM
bands
3 V/m
3 V/m
rms
rms
typical
commercial or
hospital
environment.
Electromagnetic
environment -
guidance
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the CMS70A,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
3
5 .
d
P
V
1