Alarms and Notifications
Masimo SafetyNet Alert provides visual and audible alarms when your oxygen level becomes low.
Masimo SafetyNet Alert will also contact your emergency contacts and notify if your oxygen level
drops too low and you are unaware or unable to respond.
Alarm Levels and Messages
To help ensure your safety, a notification escalation policy can be activated to establish three (3)
levels of notifications. The level of notification is based on your oxygen level and length of time at
a low oxygen level.
Level 1
Alarm Trigger
Oxygen levels are less than 85% for 30 seconds.
When Level 1 is triggered, an alarm sounds on your
Bedside Station and the App. The App also displays:
1. Your Oxygen Level in the yellow zone.
2. A WARNING banner at the top of screen.
Touch the Oxygen Level (1) or WARNING banner (2) to
learn what the Level 1 notification means.
Visual Alarm Audible Alarm Emergency Contact
Warning
Yes
(Yellow)
30
Not Contacted