Electromagnetic Immunity

Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified below. The customer or
the user of the ME Equipment should assure that it is used in such an environment.
Immunity
IEC 60601
Test
Test Level
Electrostatic
+/- 8 kV contact
discharge
+/- 15 kV air
(ESD)
IEC
61000-4-2
Electrical
+/- 2 kV for
fast
power lines
transient/
+/- 1 kV for
burst
input/ output
IEC
lines
61000-4-4
Surge
+/-1 kV line(s)
IEC
to line(s)
61000-4-5
+/- 2 kV line(s)
to earth
Conducted
3 Vrms
RF
6 Vrms
IEC
61000-4-6
Compliance Level
+/- 8 kV contact
+/- 15 air
+/- 2 kV for power lines
+/- 1 kV for input/ output
lines
+/-1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
3 Vrms
6 Vrms
61
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Performed over 0.15-80 MHz
Performed on the following ISM (industrial,
scientific and medical) bands of frequency: The
bands between 0,15 MHz and 80 MHz are
6,765 MHz to 6,795 MHz; 13,553 MHz to
13,567 MHz; 26,957 MHz to 27,283 MHz; and
40,66 MHz to 40,70 MHz. The amateur radio
bands between 0,15 MHz and 80 MHz are 1,8
MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz
to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to
10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to
18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz
to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0
MHz to 54,0 MHz