The log contents are saved when the system
is turned off. The log has limited space and
overwrites existing information when full.
You can access the log from any DICOM
configuration page.
Caution:
To avoid damaging the USB storage
device, do not remove it or turn off
the ultrasound system while the
system is exporting. In addition,
do not bump or otherwise apply
pressure to the USB storage device
while it is in a USB port on the
ultrasound system. The connector
could break.
To display the DICOM network log
1 On the Connectivity setup page, select
DICOM Setup.
2 Select Log.
To export the DICOM network log
Note: The system's Event log and the DICOM
network log have the same filename (log.txt).
Exporting either one to the same USB storage device
overwrites the existing log.txt file. For information on
the Event log, see the ultrasound system user guide.
1 Insert the USB storage device.
2 Display the DICOM network log.
3 Select Export. A list of USB devices appears.
4 Select the USB storage device, and select
Export.
Five seconds after the exporting completes, you
can safely remove the USB storage device.
To clear the DICOM network log
1 Display the DICOM network log.
2 Select Clear to delete all text.
3 Select Yes to confirm deletion.
Patient data and the worklist
The patient information form is where patient
data is entered into the system. Using the
worklist, you can import patient data from
the Hospital Information System or Radiology
Information System. (You can also enter the
data manually. See the system user guide
for instructions.)
The worklist is available after you do the
following:
• Connect the system to the network. See
"Preparation" on page 2.
•
Configure the system. See "Configuring the
system for DICOM" on page 2.
• Select a worklist server for the active location.
See "To associate devices with a location" on
page 7.
For more information about the patient
information form, see the ultrasound system
user guide.
To display the patient information form
Press the following key:
• (S Series) Patient control key
• (M‐Turbo)
PATIENT
To import patient data into the patient
information form
1 On the patient information form, select
Worklist.
The worklist appears.
2 If you want to sort the worklist, select
a column heading to sort by.
The worklist sorts in ascending order.
3 Using the touchpad and
or more procedures for the patient.
You can select multiple procedures only for
the same patient.
4 Select Select on‐screen.
key
key, select one
SELECT