Guidance and manufacture's declaration-electromagnetic emission
Guidance and manufacture's declaration –electromagnetic emission
The Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The customer of the user of the Pulse Oximeter should
assure that it is used in such an environment.
Emission
test
RF emissions
CISPR 11
RF emissions
CISPR 11
Class B
Guidance and manufacture's declaration-electromagnetic immunity
Guidance and manufacture's declaration-electromagnetic immunity
The Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The user of Pulse Oximeter should assure that it is used in
such an environment.
Immunity test
Electrostatic
discharge
(ESD) IEC
61000-4-2
Guidance and manufacture's declaration
for all EQUIPMENT and SYSTEMS
Compliance
Group 1
for all EQUIPMENT and SYSTEMS
IEC60601
test level
8kV contact
15kV air
Electromagnetic environment-guidance
The Pulse Oximeter uses RF energy only
for their internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
The Pulse Oximeter is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Compliance
level
8kV contact
15kV air
Electromagnetic
environment-
guidance
Floors should be
wood, concrete or
ceramic tile. If floor
are covered with
synthetic material,
the relative humidity