B. Please clean and disinfect the device before/after using it according to
the User Manual (6.1).
C. Please replace the batteries in time when low-battery appears.
D. Please take out the batteries if the device is not used for a long time.
E. The device need not to be calibrated during maintenance.

6.3 Transportation and Storage

A. The packed device can be transported by ordinary conveyance or
according to transport contract. During transportation, avoid strong shock,
vibration and splashing with rain or snow, and it can not be transported
mixed with toxic, harmful, corrosive material.
B. The packed device should be stored in room with no corrosive gases and
good ventilation. Temperature: -40°C~+60°C; Relative humidity: ≤95%.

7. Troubleshooting

Trouble
The SpO2 and Pulse
Rate can not be
displayed normally
The SpO2 and Pulse
Rate are not
displayed stably
Possible Reason
1. The finger is not
properly inserted.
2. The finger is
shaking or the patient
is moving.
3. The device is not
used in environment
required by the
manual.
4. The device works
abnormally
1. The finger is not
placed inside deep
enough.
2. The finger is
shaking or the patient
is moving.
Solution
1. Insert the finger
properly and
measure again.
2. Let the patient
keep calm.
3. Please use the
device in normal
environment.
4. Please contact the
after-sales.
1. Place the finger
properly and try
again.
2. Let the patient
keep calm
EN