9.2
The table below represents a guide to the potential defects of the light.
If you cannot resolve the problems, call technical support
.
Effect
The light does not
switch on
The light will flicker
The light does not
adjust light intensity
The light intensity is
considerably reduced
Stains have appeared
on the parabolas or
the reflective layer has
come away.
The light does not stay
balanced and tends to
lower
The lamp does not
control
64
GUIDE TO PROBLEMS
Cause
Power supply not
inserted or inserted
incorrectly.
Interference with
electro-surgical
scalpels or high
energy tools
Command on joystick
applied incorrectly.
Interference with
electro-surgical
scalpels or high
energy tools.
Command on joystick
applied incorrectly.
Interference with
electro-surgical
scalpels or high
energy tools.
Parabolas or
secondary lens dirty.
Use of wrong
procedures.
Use of non-approved
products.
Excessive load on
head (small mirrors,
cameras, etc..).
Synchro function off
Action (Service Engineer - SE)
Check the power supply is inserted and
the combined unit is on.
Switch off the electro-surgical scalpels
and check the permanence of the effect.
To switch on and off the light, press and
release the joystick lever using the right
or left side.
Switch off the electro-surgical scalpels
and check the permanence of the effect.
Use the command correctly as described
in this manual.
Switch off the electro-surgical scalpels
and check the permanence of the
effect.
Clean the parabolas and the secondary
lenses.
Check you have maximum adjustment
with the command.
Clean the surfaces with specific "Faro
Perflex" product. Clean the surfaces with
isopropyl alcohol.
To restore the surfaces, you need to get
service to replace the parabolas.
Remove the excess loads.
Activate the function; see 4.1.4
Resp
User
User
User
User
User
User
User
User
User
User
User