The SpO2 and Pulse Rate display
unstably
The SpO2 and PR are not
displayed normally
11. Warranty and service
We provide 1 year warranty for material and manufacturing defects of the product.
The warranty does not apply:
- in case of damage caused by improper operation
- for wearing parts
- for defects that were already known to the customer at the time of purchase
- in case of the customer's own fault
The statutory warranties of the customer remain unaffected by the guarantee.
For the assertion of a warranty claim within the warranty period, the customer must provide proof of
purchase.
The guarantee is valid for a period of 1 year from the date of purchase against
Novidion GmbH, Fuggerstr. 30, 51149 Cologne, Germany
to assert its rights. The customer has the right to have the goods repaired by us in case of a warranty
claim.
or at workshops authorised by us.
The customer is not granted any further rights based on the warranty.
The luminescent or photoelectric
window is sheltered by some
object.
Excessive movement
Finger is not placed inside deep
enough
Finger size is not within the
recommended range
Excessive ambient light
Pulse rate value of the cyclical
fluctuations
Finger is not properly positioned
Patient's SpO2 is too low to be
detected
Check the luminescent or
photoelectric window
Stop moving finger, hand and
body
Place the finger properly and try
again
Change another finger
Evite la luz excesiva
Measurement is normal, and the
patient has arrhythmia.
Place the finger properly and try
again
Try again, GO to a hospital for
diagnosis if you are sure the
device works all right.