Troubleshooting - PULOX PO-220B Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

B. Replace the batteries in time when battery voltage indicate lamps were empty.
C. It is recommended that the equipment should be kept in a dry environment with no corrosive gases
and good ventilation anytime.
D. It is best to preserve the product in a place where the temperature is between -20°C to 60°C and the
relative humidity is less than 95%.
E. The packed equipment can be tranported by ordinary conveyance. The equipment can not be
transported mixed with toxic, harmful, corrosive materials.
Warning
No modification of this equipment is allowed.
7.2 Disposal
Dispose of the pulse oximeter in accordance with local environment and waste disposal laws and
regulations.
8. Cleaning/Disinfection
Caution
Never immerse or soak the oximeter.
We recommend that the oximeter be cleaned and disinfected after use every time or determined by
your hospital's policy, to avoid long term damage to the oximeter.
Never use cleaning agents/disinfectants other than the recommended.
The sensor component is not cleaned and disinfected during testing.
8.1 Cleaning
The recommended cleaning agents include: water
A. Shut down the pulse oximeter and remove the battery.
B. Clean the oximeter with cotton or soft cloth moistened with water.
C. After cleaning, wipe off the water with a soft cloth.
D. Allow the oximeter to air dry.
8.2 Disinfection
The recommended disinfectants include: etahnol 70%, isopropanol 70%.
A. Shut down the pulse oximeter and remove the battery.
B. Clean the oximeter as instructed above.
C. Disinfect the oximeter with cotton or soft cloth moistened with one of the recommended
disinfectants.
D. After disinfection, be sure to wipe off the disinfectant left on the oximeter with a soft cloth
moistened with water.
E. Allow the oximeter to air dry.
9. Accessories
1 Lanyard, 2 AAA batteries, user manual, certificate card
Note
For particular configuration of accessories please refer to the product package list.

10. Troubleshooting

10.1 Troubleshooting
Warnings
Necessary maintenance must be performed by qualified service personal ONLY.
Users are NOT permitted to maintain the equipment by themselves.
There are NO replaceable components in the equipment.
Trouble
The Oximeter can't turn to
measure mode.
The display is off suddenly
Possible Reason
Batteries are completely
exhausted.
Incorrect battery installation
Oximeter breakdown
The device will turn into sleep
mode automatically if there is no
signal in 8 seconds.
Batteries are completely
exhausted.
Solution
Please replace batteries.
Verify and install the batteries
correctly
Please contact local service
Press the button again to
reactivate the oximeter.
Please replace batteries
EN

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido