INSTRUCTIONS FOR USE
Storage Specifications
Storage temperature
Storage Relative humidity range
Operating temperature
Operating Relative humidity range
Electrical Specifications
Not Applicable.
Software Specifications
Not Applicable.
Compatibility Specifications
Operating Room Table Compatibility
Note: Consult the corresponding IFU for the products mentioned in the above table
4.3 Sterilization Instruction:
This device is not intended to be sterilized. Equipment damage may occur.
4.4 Cleaning and Disinfection Instruction:
WARNING:
Do not use bleach or products that contain bleach to clean the Advance
Cervical Device (ACD). Injury or equipment damage can occur.
After each use, clean the Advance Cervical Device (ACD) with alcohol-based
wipes.
Do not put the ACD into water. Equipment damage can occur.
Do not use an autoclave to sterilize the ACD. Equipment damage can occur.
Use a cloth and a quaternary ammonium disinfecting/cleaning solution to clean
and disinfect the ACD.
Read and follow the manufacturer's recommendation for low-level disinfection.
Read and follow the cleaning product's instructions. Use caution in areas where
liquid can get into the mechanism.
Wipe the ACD with a clean, dry cloth.
Make sure that the ACD is dry before you store it or use it again.
Document Number: D-720774
Version: C
-29ᵒ C to +60ᵒ C
15% to 85%
This device is intended to be used in a
controlled Operating Room environment.
Not Applicable.
Not Applicable.
The Advance Cervical Device is compatible
with the following surgical table rail styles:
US, UK, EU, DEN, JP.