Aviator 3000 Series
6. Span Calibration Sequence:
Display
Lb
CAP 30
------
5 ... 0
Add20
5 ... 0
Save
Description
Calibration Unit:
kg or lb
Scale Capacity
(in the Calibration
Unit selected)
Establish zero
Countdown
2/3 calibration point
Countdown
Save or Abort
Actions
Press [FIX] to select desired unit and proceed
to the next step.
If needing to switch between kg and lb, press
[>T<].
Press [FIX] to select the displayed setting and
proceed to the next step. Make sure the
displayed value matches the scale model's
capacity.
Do not press [>T<], which will toggle through
other capacity settings that may not match the
scale model's.
Ensure the pan is empty and the scale stable.
Press [FIX] to initiate countdown sequence.
The countdown will repeat if the scale is not
stable during this process.
Press [>T<] to switch between 2/3 or full
capacity, then Place the indicated weight on
the pan, then press [FIX] to proceed to the
next step.
The countdown will repeat if the scale is not
stable during this process.
Remove the weight from the pan, then press
[FIX] to save settings and restart scale to
weighing mode.
If needing to abort calibration and retain
previous calibration data, press [>T<] ("Abort"
will be displayed) then [FIX] at this time (or turn
the scale off at any time during the process).
EN-17