EN-20

7.2 Troubleshooting

The following table lists common problems, possible causes and remedies.
If the problem persists, contact Ohaus or your authorized Ohaus dealer.
Symptom
Scale will not turn on.
Weight reading does not
stabilize.
Scale does not display
accurately.
------ Weight display
^^^^^^ Weight display
^^^^^^ Total Price display
uuuuuu Weight display
"E 11" displayed
"E 16" displayed
"E 18" displayed
"E 31" displayed
"E 33" displayed
"E 34" displayed
"noPLu" displayed
8.

TECHNICAL DATA

8.1 Specifications

The technical data is valid under the following ambient conditions:
• Indoor use only
• Operating temperature: -10º C (14º F) to 40º C (104º F)
• Humidity: Maximum relative humidity 80 % for temperatures up to 30 °C
decreasing linearly to 50 % relative humidity at 40 ° C
• Altitude: Up to 2000 m
• Mains supply voltage fluctuations: Up to ±10% of the nominal voltage
• Installation Category: II
• Pollution Degree: 2
Possible Causes
AC power not connected
to scale or to outlet.
No power in outlet.
Battery discharged.
Unstable environment.
Interference under pan or
sub-platform.
Improper calibration.
Weight exceeds power-
on zero range
Overload condition
Price > 9999.99
Underload condition
RAM error
ROM error
EEPROM error
By-Count operation error
Accumulation total price >
999.99 or the number of
transactions > 99
PLU number > 100
PLU data not found
Aviator 3000 Series
Remedy
Connect scale to power.
Connect to different power outlet.
Replace battery.
Ensure a stable environment.
Ensure that the weighing pan and
sub-platform are unobstructed.
Calibrate the scale using proper
calibration weights.
Remove excess weight from the
pan during power-on.
Reduce the load on the pan.
Reduce the load on the pan.
Install the pan.
Re-zero the scale.
Turn the scale off and on again.
Return scale for service.
Recalibrate the scale.
Return scale for service.
Return scale for service.
Enter only up to 2 digits (99)
Clear accumulation memory and
start anew
Enter correct PLU number: 1-100
Store data in the PLU number or
M-key