usIng THe sol dIode laser
Before using the SOL Diode Laser for patient treatment, it is
recommended that a sanitary, disposable single-use Barrier
Sleeve be placed over the handpiece to reduce the chance
for patient-to-patient cross-contamination. Barrier Sleeves
are SINGLE USE ONL Y. Reusing these sleeves will greatly
increase the chance of patient-to-patient infection. Safely
dispose following use. Contact your DenMat representative
for replacement packages of Barrier Sleeves.
When using the SOL Diode Laser in contact with tissue
make short quick brush strokes in light contact at the lowest
power setting that will remove the target tissue. Remove
debris that collects on the fiber tip with a clean 2 x 2 gauze
sponge moistened with water. DO NOT USE ALCOHOL OR
OTHER FLAMMABLE LIQUIDS TO WET THE SPONGE.
When the laser is not in use and when the procedure has
been completed, press the Ready Key to return to the
Stand-By Mode.
Turn the Key Switch counterclockwise to the "OFF" position
remove and store in a safe and secure location when the
laser is not in use.
Turn the rear Power Switch to the "OFF" position
to conserve battery power.
Record the Power settings and total procedure times used
for each procedure in the patient's chart.
j ThE DEnmAT hOlDings, llC
sysTEm sPECiFiCATiOn
THe sol laser Is Class II
SOL Laser relies upon having two layers of insulation
surrounding live parts or by having reinforced insulation to
prevent electric shock.
THe sol laser Is TyPe bf
BF
As type B but with isolated or floating (F – type)
applied part or parts.
ClassIfICaTIon agaInsT flammable
vaPors WITH aIr
SOL Laser is NOT suitable for use in the the presence of
Flammable Anaesthetic Mixture with Air, Oxygen or
Nitrous Oxide.
10
10
k sOl lAsER sPECiFiCATiOns
Laser Classification: Laser diode Class 4 laser device
(per IEC 60825-1:2007)
Wavelength working beam: 808 nm +/- 5 nm
Beam divergence: 9° +/- 1°
Power range: 100mW to 3.0 Watts CW. 100mW to
5 Watts Pulse
Frequency (Pulsed): Fixed, 10 Hz
Continuous Wave (CW) Duty Cycle: Fixed 100%
Pulse Mode Duty Cycle: Fixed 50%
Aiming Beam Output: 2mW Maximum, adjustable
Audible signal: Yes, adjustable On/Off
Visual signal: Yes, adjustable
Power Supply: 30 Watts +12VDC supplied from
100-240VAC 50-60 Hz MODEL: FW7362M/12
Weight: 1.16 Kg
ComPlIes WITH
IEC 60601-1, IEC 60601-2-22; IEC 60825-1:2007; IEC
60601-1-2, Class B Per EN 60601-1-2 as well as to US
Federal Regulations for laser performance standards set
out in 21CFR 1040.10 and 1040.11 except for deviations
pursuant to laser notice 50 dated July 26, 2001.
Hand PIeCe
Length: 12.1 cm
Diameter: 1.50 cm
Tip removal: Manual friction-fit
Sanitizing: Wipe with appropriate disinfecting solution
fIber oPTIC TIPs
Type: Removable, Single-Use Disposable
Material: Fused Silica, coated
Size: 400 μm
Style: Polymer-clad fiber
aCCessorIes
Sanitary, disposable single-use only Barrier Sleeve for
Handpiece
400 μm Single-Use ONLY Fiber Optic Tips
Contact DenMat Sales to order accessories.
l sOl sERviCE AnD
TROUblEshOOTing
Recalibration is recommended at a minimum
of one time per year based on average usage
(see Calibration page
9).