TROUBLE-SHOOTING
TYPE OF FAULT DISPLAYED, CONTINUOUS ALARM SOUNDS:
OPERATOR ACTION REQUIRED
First stop the alarm by pressing the key
Depending on the type of fault displayed, consult the list below to determine the cause or causes
of failure and contact your After-Sales Service.
The different types of faults are numbered 1 to 63 (broken down by powers of 2)
FAILURE 1: THERMOCOUPLE MALFUNCTION
The thermocouple is broken.
FAILURE 2: HEAT MALFUNCTION
The temperature exceeds the programmed final temperature by + 50° C.
This malfunction is generally caused by an anomaly in TRIAC 6 on the furnace circuit board.
FAILURE 4: VACUUM PUMP MALFUNCTION
There are two main causes:
* The pump is out of order
* The electric or pneumatic connection is defective or non-existent.
FAILURE 8: AIR MALFUNCTION
The solenoid valve is clogged or improperly connected to the board (GREEN/GREEN WIRE).
FAILURE 16-32: PLATFORM MALFUNCTION (UP-DOWN)
The platform is blocked in position, for three possible reasons:
* The back-geared motor is out of order
* The transmission belt is broken
* The electric connection on the board is defective (RED/BLUE WIRE).
* After the correction of the problem, push the button
«Start the cycle again for verification (just for a test).
CAUTION: The list given above is only partial, in that certain displays may combine several
failures:
Example: FAILURE 22
Whatever the case, it is essential that you contact your After-Sales Service.
FAILURE 16+4+2
in order to restore the normal display
33
ARTIS